ZEN50 User Guide September 2015 – Rev 2
Page 14
Self Test Procedure
Each time the Atlas ZEN is powered up, a self test procedure is performed. In
addition to a battery voltage test, the unit measures the performance of many
internal functions such as the voltage and current sources, amplifiers and the
analogue to digital converters. If any of these function measurements fall
outside tight performance limits, a message will be displayed and the
instrument will switch off automatically.
If the problem was caused by a temporary condition on the test clips, such as
applying power to the test clips, then
simply re-starting the Atlas ZEN may
clear the problem.
If a persistent problem does arise, it is likely that damage has been caused by
an external event such as excessive power being applied to the test clips or a
large static discharge taking place. If the problem persists, please contact us for
further advice, quoting the displayed fault code.
If there is a low battery condition, the automatic self test procedure
may not be performed. For this reason, it is highly recommended that
the battery is replaced as soon as possible following a “Low Battery”
warning.
LCD Contrast Adjustment
The contrast of the LCD can be adjusted to your personal preference or
viewing angle. There are 6 steps of contrast adjustment available.
To step through each setting, simply hold down the on-test button and briefly
press the scroll button repeatedly until your desired contrast is obtained. Then
release the on-test button.
Self test failed
CODE: 5